SETUP FOR SIMULATION

ReMoving and ReplaCing the iv aRM pad

3 Remove excess blood that may accumulate in the IV

Arm chassis.

4 When replacing the IV arm pad, ensure that it is pressed

firmly into place.

1 Ensure that the Tourniquet Area is not activated i.e.

remove the tourniquet.

2 Remove the IV Arm Pad by pulling at the tabs on edge

of the IV pad.

5 Fill and Prime the system as instructed in the previous section

Refilling the iv aRM with Blood

1 If the veins do not respond (energize) when the

tourniquet is applied it is likely that the system requires
refilling with blood.

2 Connect a blood filled syringe and slowly fill with no

more than 16ml of blood or until you feel resistance.
Remove the Syringe.

The IV Arm is now ready for use.