TROUBLESHOOTING

79

• Check the flow setting in the Instructor Application

• Replace filter

• The blood system may need to be re-calibrated. Contact

your local service representative.

Problem

• Air when bleeding

Solution

• Bleed the blood reservoir empty and refill.

Problem

• Simulated blood is seen leaking from the back of the right

leg during filling or when the manikin is turned off.

Solution

• Blood system has been overfilled, when running manikin

with external blood. Empty internal reservoir and fill
according to instructions.

• The internal reservoir may need to be replaced. Contact

your local service representative.

Problem

• Can’t fill the blood reservoir

Solution

• The filter in the fill unit may be clogged. If so, order a

replacement fill unit.

Manikin liMBS

Problems

• Lack of motion in the legs

Possible Solution

• Loosen and re-adjust the hip joint nuts on the inside of the

pelvis. See section: Manikin Setup - Attaching the legs.

lunGS

Problem

• Lungs not functioning properly

Possible Solution

• Check that the Airway resistance is not set to maximum in

the Instructor Application.

• Open the torso and chest plate. Check that the lungs are

free to expand and are not restricted by any cables.

• Check that the lung bladders are properly connected, and

that the tubes are not twisted.

• Check that the lung bladders are in a horizontal position

and inserted correctly. Ensure that the lung compliance
O-rings lie between the folds of the lung bladders.

• Check for flaws or ruptures in the lung bladders.

• Check that the two lung compliance O-rings are fitted

correctly. Replace O-rings if they appear to be damaged.

• Check that there are no obstructions inside the manikin

airways which may block air flow.

• If there is no change when adjusting lung compliance,

contact Laerdal Technical Service.

• If there is no change in lung resistance, contact Laerdal

Technical Service.

Mechanical noiSe durinG auScultation
In the Instructor Application, click <auscultation focus>.

BatterieS

Problem

• Battery life is less than 150 minutes with healthy patient

and fully charged batteries.

Possible Solutions

• Batteries may be old, (recommended lifetime is 200

discharge cycles). Insert new batteries.

• Compressor may not function properly; consult your local

Laerdal Technical Services Centre.

PulSeS

Problem

• Cannot feel pedal pulses

Possible Solutions

• Skin may be too tight over pulse units – re-adjust skin and

reboot.

clinical featureS

-

PneuMothorax BladderS

Problem

• Pneumothorax bladders are not functioning properly.

Solution

• Check tubing connection at base of the bladder, to ensure

that tubing has not been disconnected.

Manikin ShutdoWn

Problem

• The manikin is unresponsive.

Solution

• Press and hold the <ON/OFF> button for 10 sec to force

shutdown the manikin.

urine SySteM

Problem

• No urine output after filling the urine system.

Solution

• Make sure that the manikin is laying on its back, the urine

tank is placed correctly in the pelvis and no catheter is
inserted. Fill the urine system with 100ml of air and then
insert a catheter and drain the urine system. Remove
catheter. Repeat until all the fluid inside the urine tank has
drained out.

Note: Do not fill the urine system with the catheter inserted.