Changing the irises – Laerdal SimMan Essential User Manual

Page 57

Advertising
background image

57

SETUP FOR SIMULATION

Changing the Irises

The irises can be changed to simulate different clinical states. The irises are secured by magnets in the eye socket. To change the irises
use the especially adapted suction wand.

Before changing the irises during a simulation, stop the blinking function in the Instructor Application.

1 Moisten the suction cup before performing this procedure.

1 With one hand pull back the upper lid.

2 Place the suction wand on the upper part of the iris.

Squeeze the back end of the wand to create a suction
seal on the iris.

3 Pull and roll the iris up and down towards the manikin’s mouth

using a rolling motion.

4 Place the iris in the eye kit and select another from the kit.

5 Insert the lower part of the iris into the socket until it connects

with the magnet below. Repeat this for the other eye.

SimMan Ess DFU_RevF.indd 57

18.07.12 14:08