Ks s φ – Bronkhorst IN-FLOW User Manual

Page 18

Advertising
background image

BRONKHORST HIGH-TECH B.V.

page 18

9.17.022

The approximate accuracy of the conversion factors listed is:

typical for conversion factors;

> 1

2% x factor

< 1

2% / factor

However, as the accuracy of the factor also depends on viscosity, pressure and temperature, special

attention should be taken for gases in the gas/liquid state where specific heat, density and viscosity can vary

tremendously. Apply to factory for more detailed information.

For gas mixtures a good approach is the following simplified equation:

1

1

1

2

2

C

V

C

V

C

mix

=

+

+

.....

V

C

n

n

C

mix

= Conversion factor for gas mixture

C

n

= Conversion factor for gas n

V

n

= Volumetric part of gas n in the mixture

Example Gas mixture contains:

(1) 10% N

2

C1 = 1,00

(2) 30% Ar

C2 = 1,40

(3) 50% CH

4

C3 = 0,76

(4) 10% He

C4 = 1,41

1

010

100

0 30

140

0 50
0 76

010

141

1043

C

mix

=

+

+

+

=

,

,

,

,

,
,

,

,

,

C

mix

= 0,959

When the original meter has been calibrated on 500 ml

n

/min N

2

, 100% means:

500

00

,1

959

,

0

= 480 ml

n

/min mixture.

When the original meter has been calibrated on 500 ml

n

/min Argon, then 100% means:

500

40

,1

959

,

0

= 343 ml

n

/min gas mixture.

1.6.2 Gas Conversion Factors (direct mass flow measurement, CTA-based)

For CTA-based gas flow sensors the general relationship between signal and mass flow is:

n

m

signal

K

S

S

Φ

+

0

In which:

S

signal

= output signal

S

0

= offset (zero flow) signal

K

= constant factor (includes λ – heat conductivity, C

p

– specific heat, μ – dynamic viscosity

and ρ – density of the gas)

Φ

m

= mass flow

n

= dimensionless constant (typically of order 0.5)

Due to the offset signal (which is also dependent on fluid properties) and the non-linear relationship between

signal and mass flow, a single conversion factor for a custom fluid that covers the entire flow range of an

instrument can not be obtained. However, a complex and partially empirical conversion model is available for

most common gases, which is accurate at both lower and higher flow ranges. Consult Bronkhorst High-Tech

B.V. for applications.

Advertising