8 analog operation, 9 bus / digital operation – Bronkhorst IN-FLOW User Manual

Page 27

Advertising
background image

BRONKHORST HIGH-TECH B.V.

9.17.022

page 27

0V common

-

Current output signals

Sourcing

instrument

output

I

+

mA

3.8

Analog operation

Digital instruments can be operated with analog signals through the 9-pin sub-D connector or 8DIN

connector. The instruments are compatible in use with

analog instruments on this point.

Analog operated instruments can be hooked-up using an 8-wire shielded cable with 9-pin D-connectors or

8DIN connectors, connected according to the Bronkhorst High-Tech standard.

Each electronic p.c.board is set for one of the following output (and corresponding input) signals:

Signal

output (sensor)

input (setpoint)

code

signal

signal

A

0…5 Vdc

0…5 Vdc

B

0…10 Vdc

0…10 Vdc

F

0…20 mA (sourcing)

0…20 mA (sinking)

G

4…20 mA (sourcing)

4…20 mA (sinking)

For meters only the output signal is available.

At analog operation following parameters are available:

- measured value

- setpoint (controllers only)

- valve voltage (controllers only)

Note:

When operating the instrument through the analog interface it is possible to connect the instrument to any

supported fieldbus system (or RS232-interface with special cable) for reading/changing parameters (e.g.

controller response or other fluid selection).

For FLOW-BUS versions of the instruments a readout/control module for digital instruments can be

temporarily connected to the RJ45 modular jack plug.

3.9

BUS / digital operation

Operation via fieldbus reduces the amount of cables to build a system of several instruments and offers more

parameter values to be monitored/changed by the user.

See instruction manual: operating digital mass flow / pressure instruments for more details (document nr.

9.17.023).

Operation by means of a fieldbus adds a lot of extra features (compared to analog operation) to the

instruments.

Such as:

- setpoint slope (ramp function on setpoint for smooth control)

- 8 selectable fluids (calibration settings for high accuracy)

- direct reading at readout/control module or host computer

- testing and self diagnosis

- response alarm (|setpoint-measure| too high for too long time)

- several control/setpoint modes (e.g. purge/close valve)

- master/slave modes for ratio control (FLOW-BUS only)

- identification (serialnumber, modelnumber, device type, user tag)

- adjustable minimal and maximal alarm limits

- (batch) counter

- adjustable response time for controller when opening from zero

- adjustable response time for normal control

- adjustable response time for stable control (|setpoint-measure| < 2%)

Advertising