10 supply pressure, 11 system purging, 12 seals – Bronkhorst IN-FLOW User Manual

Page 23: 13 equipment storage

Advertising
background image

BRONKHORST HIGH-TECH B.V.

9.17.022

page 23

Fold the shield of the cable back over the cable (the shield must be around the cable).

Wind a copper tape around the shield

Solder a black wire on the tape and

connect to pin 9 of the connector.

copper tape

shielded cable

e.g. LAPP LiYCY

other wires

D-connector housing

metalized.

connector

8 mm

20 mm

Black wire

(shield)

2.10

Supply pressure

Do not apply pressure until electrical connections are made. When applying pressure to the system, take

care to avoid pressure shocks in the system and increase pressure gradually, especially on high pressure

units incorporating a piston operated control valve.

2.11

System purging

If explosive gases are to be used, purge the process with inert dry gas like Nitrogen, Argon etc. for at

least 30 minutes.

In systems with corrosive or reactive fluids, purging with an inert gas is absolutely necessary, because if the

tubing has been exposed to air, introducing these fluids will tend to clog up or corrode the system due to a

chemical reaction with oxygen or moist air.

Complete purging is also required to remove such fluids from the system before exposing the system to air. It

is preferred not to expose the system to air, when working with these corrosive fluids.

2.12

Seals

Bronkhorst High-Tech B.V. has gathered a material compatibility chart from a number of sources believed to

be reliable.

However, it is a general guide only. Operating conditions may substantially change the accuracy of this guide.

Therefore there is no liability for damages accruing from the use of this guide.

The customers application will demand its own specific design or test evaluation for optimum reliability.

So check if the seals like O-rings, plunger and packing gland of capillary are correct for the process.

2.13

Equipment storage

The equipment should be stored in its original packing in a cupboard warehouse or similar. Care should be

taken not to subject the equipment to excessive temperatures or humidity.

2.14

Electromagnetic compatibility

2.14.1 Conditions for compliance with EMC requirements

All instruments described in this manual carry the CE-mark.

Therefore they have to comply with the EMC requirements as are valid for these instruments.

However compliance with the EMC requirements is not possible without the use of proper cables and

connector/gland assemblies.

For good results Bronkhorst High-Tech B.V. can provide standard cables. Otherwise follow the guidelines as

stated below.

D-Connector assembly

Advertising