Bronkhorst Multibus User Manual

Page 31

Advertising
background image

BRONKHORST

®

Page 31

Operational instructions for digital multibus instruments

9.17.023

7.4 C

OUNTER NEW SETPOINT

Unsigned int

0…32000

RW

N

127

104/6

New/safe setpoint at counter limit/batch situation (until reset). See measure for range.

Normally this value is set to 0%.

7.5 C

OUNTER LIMIT

float

0…9999999

RW

N

124

104/3

Counter limit/batch in units selected at Counter unit. Value is a float in IEEE-754 32-bits single precision notation. Default

setting is 0 ln.

7.6 C

OUNTER UNIT INDEX

Unsigned char

0…13

RW

N

123

104/2

This parameter gives access to the limited unit table which is available for MBC-II and MBC3 type of

instruments.

Counter unit index is a pointer to select an actual readout unit (see list below).

counter unit index table (limited unit table)

0

1

2

3

4

5

6

7

8

9

10

11

12

13

Sensor

Type

1

l

mm3

ml

cm3

ul

m3

2

g

mg

ug

kg

3

ln

mm3n mln cm3n

uln

dm3n m3n

uls

mm3s mls

cm3s

ls

dm3s m3s

Sensor type number explanation:

nr

Sensor type

0

pressure (no counting allowed)

1

liquid volume

2

liquid/gas mass

3

gas volume

4

other sensor type (no counting allowed)