4 operating instructions – Fluke Biomedical 35080B User Manual

Page 22

Advertising
background image

35080B
Operators Manual


4-2

4.4 Operating Instructions

The operation of the Test Filter Pack (32863) is similar to the operation of the Wide Range Filter Pack
(37617) described in Section 2. Prior to taking measurements, it is suggested that the battery and
internal calibration be checked as described in Sections 2.1 and 2.2. Position the kVp Divider as
described in Section 2.5, Geometry of Setup.

Set up the oscilloscope to read the 3.0 volts expected from the Model 35080B kVp divider using the Test
Filter Pack (32863), and make an exposure. On a properly functioning unit, the observed waveform
amplitude will be 3.00 Volts ± 1%. Be aware that the oscilloscope accuracy may not be adequate for this
test. The calibration of the kVp Divider can only be verified to within the accuracy of the oscilloscope or
other readout device used. If the unit appears to be out of calibration, refer to Section 10.4, Calibration.




















Figure 4-1.

The Model 35080B kVp Divider Single-phase and Three-phase Output with the Test

Filter Pack (32863)

kV WaveForm
Single-Phase

kV WaveForm
Three-Phase

Model 35080B WaveForm
Output Single-Phase

Model 35080B WaveForm
Output Three-Phase

Advertising