Linear Mammo Filter Pack Option (37351)

Operating Instructions

6


6-5

5. Measure other kVp values as desired. Record the LINEAR VOLTAGE, multiply by 100 and this will

be your uncorrected LINEAR KVP. Lastly, add or subtract the CORRECTION VALUE and record
the results. This now has the true kVp value for all kVp's measured.



See Appendix A for step-by-step instructions on
using the Linear Mammo Filter Pack (37351) in
conjunction with the Cadmium K-edge Mammo
Filter Pack (37355).

6. Subtract this kVp voltage from the true K-edge kVp value (see Section 5.4). (This will be the

CORRECTION VALUE applied to all kVp measurements made with the Linear Mammo Filter Pack,
37351.)