2 specifications, 3 theory of operation – Fluke Biomedical 35080B User Manual

Page 24

Advertising
background image

35080B
Operators Manual


5-2

5.2 Specifications

Intended Use

Beryllium window, molybdenum anode mammographic generators

Range

27.5 kVp to 29.5 kVp

Accuracy

±0.5 kV

Sensitivity

0.1 volt per kV

Zero

27.0 kV

Cal

0.75 V

Minimum Anode Current

50 mA

Target to kVp Divider Distance

12 in (30.5 cm)

Effect of Molybdenum Filtration

Negligible, 12 to 35 microns

5.3 Theory of Operation

The Cadmium K-edge Mammo Filter Pack (37355) contains two different filters side-by-side. One is
constructed of cadmium, an element whose x-ray attenuation increases in a step-fashion at its K-edge
(26.7 kV). The other filter is made of an element like aluminum that has no K-edge in the mammographic
region. These two filters are selected to have identical attenuation characteristics for photon energies
less than 26.7 kV.

The Model 35080B kVp Divider takes the ratio of the x-ray signals passing through the two filters. As long
as the applied voltage to the x-ray generator is less than 26.7 kV, the output of the kVp Divider is constant
(approximately 3 volts). When the applied voltage exceeds 26.7 kV, the ratio and thus the output voltage
increases rapidly. The filters are carefully manufactured so that the voltage increases at 0.1 V/kV above
27.0 kV. For molybdenum anode, beryllium window generators, there is very little effect on the calibration
for changes in molybdenum filtration from 12 to 35 microns. As shown by Figure 5-2, the addition of
aluminum filtration does influence the sensitivity. Adding 0.25 mm of aluminum can increase the kVp
divider's reading by up to 0.5 kVp.

Advertising