4 operating instructions – Fluke Biomedical 35080B User Manual
Page 25

Cadmium K-edge Mammo Filter Pack Option (37355)
Theory of Operation
5
5-3
Figure 5-2.
The kVp Correction Curves for the Cadmium K-edge Mammo Filter Pack (37355)
5.4 Operating Instructions
1. Measurement Setup - Place the Model 35080B kVp Divider in the x-ray beam on the compression
paddle or other suitable support with the axis of the kVp Divider perpendicular to the axis of the x-
ray tube. Raise the kVp Divider (with the Cadmium K-edge Mammo Filter Pack, 37355) as close as
possible to the x-ray tube. The x-ray beam should cover the entire filter pack. (See Figure 5-3.)
Figure 5-3. K-edge Measurement Setup
Minimum
Model 35080B
kVp Divider
kVp C
o
rrection
Indicated kVp
Cadmium K-edge Mammo Filter
Pack (37355) kVp Correction Curve