Cadmium K-edge Mammo Filter Pack Option (37355)

Theory of Operation

5


5-3






















Figure 5-2.

The kVp Correction Curves for the Cadmium K-edge Mammo Filter Pack (37355)

5.4 Operating Instructions

1. Measurement Setup - Place the Model 35080B kVp Divider in the x-ray beam on the compression

paddle or other suitable support with the axis of the kVp Divider perpendicular to the axis of the x-
ray tube. Raise the kVp Divider (with the Cadmium K-edge Mammo Filter Pack, 37355) as close as
possible to the x-ray tube. The x-ray beam should cover the entire filter pack. (See Figure 5-3.)

















Figure 5-3. K-edge Measurement Setup

Minimum

Model 35080B

kVp Divider

kVp C

o

rrection

Indicated kVp

Cadmium K-edge Mammo Filter
Pack (37355) kVp Correction Curve