Fluke Biomedical 35080B User Manual

Page 53

Advertising
background image

Maintenance

Calibration

10

10-5


If the oscilloscope in use is not accurate to ± 0.1%,
do not attempt adjustment of the Model 35080B
kVp Divider.

OTHER ADJUSTMENTS

The third adjustment potentiometer (RP103) adjusts the amount of the waveform that is displayed by the
oscilloscope. If annoying artifacts are observed on the oscilloscope, turning the potentiometer (RP103)
counterclockwise will decrease the width of the waveform displayed on the oscilloscope. Turning the
same potentiometer clockwise will conversely increase the width of waveform displayed. Refer to Figure
10-3 for an illustration of the effects of this adjustment.



























Figure 10-2.

Location of RP101, RP102, and RP103

NOTE

Advertising