Fluke Biomedical 35080B User Manual

Page 9

Advertising
background image

General Information

Specifications

1

1-3

Table 1-2.

The Wide Range Filter Pack (37617) Specifications

Range: 50 to 150 kVp

Sensitivity: 0.1 V/10 kV

Cal: 0.400 V

Zero: 40 kV

Cal: 80 kV

Accuracy: ± 2% exclusive of linearity and filtration
effects
Linearity and filtration effects can be corrected using
curves from Section 2 of this instruction manual.

Optional Filter Packs

A Test Filter Pack (32863) that provides verification of instrument accuracy, a K-edge Mammo Filter Pack
(37355)* and a Linear Mammo Filter Pack (37351) used for precision measurements from 22 to 40 kVp, a
CT Filter Pack (33551) used for CT system kVp measurements from 70 to 140 kVp, a Mobile Filter Pack
Plus (37946) used for measurements on x-ray systems with kV waveforms above 1 kHz with substantial
ripple, and a Low Range Filter Pack (38237) used for measurements from 30 kVp to 90 kVp where the
kVp range or sensitivity of other filter packs is not adequate.

* Patent numbers 4,843,619 and 4,916,727