Operators Manual

9-4
























Figure 9-2. Low Range Filter Pack (38237) Linearity Correction Curve. The kVp Correction Value read

from the vertical scale should be added to the Model 35080B Model 35080B kVp Divider
reading to obtain the corrected kVp result.

Low Range Filter Pack (38237)

Linearity Correction Curve

-1.5

-1.0

-0.5

0.0

0.5

1.0

1.5

2.0

20

30

40

50

60

70

80

90

Model 35080B kVp Reading

kVp Correction