Fluke Biomedical 35080B User Manual
Page 63
Advertising

Appendix
The Model 35050A Dosimeter Measurements
B
B-3
(
)
offset
Zero
offset
Zero
kVp
CAL
Actual
offset
Zero
kVp
CAL
Expected
offset
Zero
kVp
Measured
kVp
Corrected
+
−
−
−
=
⎟
⎠
⎞
⎜
⎝
⎛
Example: For the Wide Range Filter Pack (37617) (50-150kVp), the Expected CAL kVp is given as 80,
and the Zero offset value is 40 kVp. However, a Model 35080B kVp Divider in CAL mode with this filter
pack inserted is connected to a Model 35050A Dosimeter, and a kVp of 81.2 is measured. When placed
in an x-ray beam in RUN mode, the Model 35050A Dosimeter reads 91.5 kVp. The Corrected kVp for this
exposure is calculated from the previous equation.
(
)
kVp
90
=
kVp
40
kVp
40
-
kVp
81.2
kVp
40
-
kVp
80
kVp
40
-
kVp
91.5
=
kVp
Corrected
+
⎟
⎠
⎞
⎜
⎝
⎛
Advertising