Introduction

9

9-1

Section 9

Low Range Filter Pack Option (38237)


9.1 Introduction

The Low Range Filter Pack (38237) is optimized for measurements in the 30-kVp to 90-kVp range. It can
be used to make measurements below the 50-kVp lower limit of the Wide Range Filter Pack (37617) and
Mobile Filter Pack Plus (37946) on x-ray generators with settings below 50 kVp. The Low Range Filter
Pack (38237) will typically operate at lower mA settings, making it useful in taking measurements difficult
to obtain with either the Wide Range Filter Pack (37617) or Mobile Filter Pack Plus (38237) at the lower
end of their kVp operating range.

9.2 Specifications

Range

30 kVp to 90 kVp

Sensitivity

0.1 V/10 kV

Cal

0.150 V

Zero

26 kV

Cal

41 kV

Accuracy

± 2%, exclusive of linearity and filtration effects

Minimum Anode Current for less than ± 1 kV effect

Typical at 22 inches with a 3 phase x-ray generator, 3 mA from 90 kVp down to 50 kVp, increasing to
10 mA @ 40 kVp and 30 mA @ 30 kVp.

9.3 Theory of Operation

The Low Range Filter Pack (38237) operates in a manner similar to the operation of the Wide Range
Filter Pack (37617) described in Section 3. The thickness and composition of the filters have been
optimized for operation over the 30 kVp to 90-kVp range in the Low Range Filter Pack (38237). The filter
differences cause the different Cal and Zero voltages specified, and require different correction curves
shown later in this section.