Specifications, Electrical specification, Flow rate measurement – Fluke Biomedical IDA-4 Plus User Manual

Page 15: Volume measurement, Pca bolus measurement, Pressure measurement

Advertising
background image

Infusion Pump Analyzer

Introduction

1

1-3

Specifications

Electrical Specification:

Supply Voltage

90 - 260 VAC.

Supply Frequency

50 - 60 Hz.

Supply Power

< 30 VA.

Fuse

20 mm 250 V, 1 A (T) (slow blow).

Earth Leakage Current

< 1.0 mA in single fault condition.

Flow Rate Measurement

Technique: Flow is calculated by measuring a volume over time.

Range:

0.5 - 1000 ml/hr.

Accuracy:

1% of reading

± 1 LSD for flows of 16 - 200 ml/hr for volumes over 20 ml.

Otherwise, 2% of reading

± 1LSD after delivery of 10 ml under laboratory

conditions.

Volume Measurement

Technique: Volume is measured directly by the transducer in minimum sample sizes of

60 micro-liters.

Range:

0.06 - 9999 ml.

Accuracy:

1% of reading

± 1 LSD for flows of 16 - 200 ml/hr for volumes over 20 ml.

Otherwise, 2% of reading

± 1LSD after delivery of 10 ml under laboratory

conditions.

PCA Bolus Measurement

Technique: Volume is measured directly by the transducer in minimum bolus volumes

of 0.5 ml. The measurement is made with a continuous rate between 0.0 and
30 ml/hr. The bolus flow rate should be at least four times the basal flow
rate for reliable detection of boluses.

Min Bolus: 0.5 ml.

Accuracy:

See volume measurement.

Pressure Measurement

Technique: Direct occlusion of the infusion line and measurement of pressure prior to

the glass transducer.

Range:

0 to 45 PSI and equivalents in mmHg and kPa.

Accuracy: 1% of Full Scale

± 1 LSD under laboratory conditions.