PCA Activity with Basal Flow

This will follow the PCA Activity Only scenario
described above, but with a basal flow rate set. The
basal flow rate should also be entered on the
IDA-4 Plus.

Basal Flow Only

When a PCA pump is operated with the basal flow rate
only, it should be tested using the normal single flow
rate method.

Optional PCA Trigger / Remote Call Interface

CHANNEL 1

READY

CH1

CH2

PCA TRIGGER OUTPUT

REMOTE CALL INPUT

IDA4 Plus

PCA TRIGGER/REMOTE CALL INTERFACE

CH3

CH4

CH1

CH2

CH3

CH4

TRIGGER

CHANNEL 2

READY

CHANNEL 3

READY

CHANNEL 4

READY

TRIGGER

TRIGGER

TRIGGER

The PCA Trigger / Remote Call Interface allows PCA
pumps to be triggered by the IDA-4 Plus. It also has
connections to allow alarm conditions of pumps to be
monitored. It connects to the IDA-4 Plus via the 15
Way socket on the rear panel.

Trigger

One trigger button is provided for each channel, to
trigger a PCA pump via the PCA trigger output. This
eliminates the need to ensure that the IDA-4 Plus is
displaying the correct screen.

Ready

This LED will illuminate indicating that a bolus can be
triggered on that channel. When a bolus is detected the
LED will extinguish.

PCA Trigger Output

The PCA Trigger Output jacks (one for each channel)
provide floating relay contacts. Normally open contacts
are available between ring and shield. Normally closed
contacts are available between tip and shield. (Refer to
the pump manufacturer’s service manual when
preparing a PCA trigger cable).

from interface box

1/4” phono

shield

PCA trigger output configuration

normally

open

contacts

normally

closed

contacts

to PCA pump

tip

rin

g

gn

d