Occlusion test screen (start mode), Occlusion test screen (measuring mode, Pressure – Fluke Biomedical IDA-4 Plus User Manual

Page 42

Advertising
background image

IDA-4 Plus
Operators Manual

3-20

Occlusion Test Screen (Start Mode)

When the transducer has completed its tare operation
the screen changes as shown.

Start

When the transducer has completed its tare
measurement the START option is enabled. Pressing
ENTER while this option is selected will start the
occlusion test and advances to the measuring mode.

Status

This invokes the STATUS ALL CHANNELS screen.

ESC

Selecting ESC or pressing the ESC key on the external
keyboard will take the operator back to the CHANNEL
SETUP
menu.

Occlusion Test Screen (Measuring Mode

When the test is started the screen enters the measuring
mode as shown. This enables further menu options.

Status

This invokes the STATUS ALL CHANNELS screen.

Graph

This switches to the OCCLUSION GRAPH screen for
the current channel.

Flow

This ends the current test, clears the values from the
screen and proceeds to the FLOW TEST screen in
either PRIME or START mode, depending on the
current state of the transducer. The results of the
subsequent test will be appended to those of the test
terminated by this option.

Occlusion

This ends the current test, clears the values from the
screen and proceeds to the OCCLUSION TEST screen
in Wait Mode. The results of the subsequent test will be
appended to those of the test terminated by this option.

End

Ends the test, or sequence of tests, and advances the
OCCLUSION TEST screen to the End of Test Mode.

C H A N N E L 1 O C C L U S I O N

0.00

PRESSURE

psi

TIME

00 : 00 : 00

Peak Pressure

psi

Time of Peak 00: 00: 00

START

STATUS ESC

C H A N N E L 1 O C C L U S I O N

0.00

PRESSURE

psi

TIME

00 : 00 : 00

Peak Pressure

psi

Time of Peak 00: 00: 00

STATUS

GRAPH FLOW OCCL END