Pca test screen, Pca test screen (prime mode), Prime – Fluke Biomedical IDA-4 Plus User Manual

Page 46

Advertising
background image

IDA-4 Plus
Operators Manual

3-24

bolus will be reported as a loading dose, otherwise a
loading dose will not be reported.

On completion of this field, the PCA TEST screen is
invoked.

Note: When testing PCA pump performance it is
essential that new syringes be used. It has been
observed that syringe performance degrades rapidly
after about 10 uses. The deterioration is particularly
noticeable at the start of a bolus when the flow rate is
increasing rapidly. It is characterized by a very high
initial flow rate at the start of the bolus (4 to 5 times the
bolus flow rate). This can cause the end of the bolus to
be detected at the end of the spike.

PCA Test Screen

This screen displays the measured output of the flow
rate, volume and duration of the last bolus delivered,
and the average flow rate, volume and duration of all
boluses delivered. It will also display the measured
basal rate, total delivered volume, total test time and
time interval between the last two boluses delivered.

PCA Test Screen (Prime Mode)

When the screen is first displayed, the operator is
prompted to prime the channel. This is essential to
ensure that the measuring circuit is filled with fluid. It is
particularly important that the instrument is carefully
primed when the loading dose is to be measured.

Prime

This prompt stays on the screen while there is any air in
the measuring circuit. Pressing ENTER while this
prompt displays PRIME will not have any effect. As
the circuit is primed the prompt will change to START
and advance to Start Mode. There may be some
flickering between PRIME and START as bubbles are
pushed through the system.

ESC

Selecting ESC or pressing ESC on the external
keyboard takes the operator back to the CHANNEL
SETUP
menu.

C h a n n e l 1 P C A

L a s t B o l u s

A v e r a g e

0 . 0 0

V o l u m e m l

0 . 0 0

0 . 0 0 F l o w

R a t e

0 . 0 0

0

D u r a t i o n

0

I n t e r v a l

B o l u s C o u n t

0

B a s a l F l o w R a t e

0 . 0 0 m l / h

E l a p s e d T i m e

0 0 : 0 0 : 0 0

T o t a l V o l u m e

0 . 0 0 m l

PRIME

STATUS ESC