Fluke Biomedical IDA-4 Plus User Manual
Page 74

IDA-4 Plus
Operators Manual
B-6
ASCII Code
Function
Number of
Parameters
IDA-4 Plus
< - - >
BYE Terminates
Computer Control
0
- - >
POLL
Returns # of
channels installed
0
< - - >
PRESUn
Sets the pressure
units
n = Channel number
- - >
CnF,a,b,c
Starts a Flow Test
3,
n = Channel number
- - >
CnV,a,b,c
Starts a Volume
Test
3,
n = Channel number
- - >
CnO,a,b,c
Starts an Occlusion
Test
3,
n = Channel number
- - >
CnPCA,a,b,c
Starts a PCA test
3,
n = Channel number
- - >
CnD,a,b,c,d,e,f
Starts a Dual Flow
Test
6,
n = Channel number
- - >
END,n
End a Test
n = Channel number
- - >
FLOWn Fetches
current
flow
rate
n = Channel number
< - - >
VOLn Fetches
current
volume
n = Channel number
< - - >
PRESn Fetches
current
pressure
n = Channel number
< - - >
PKPRESn Fetches
the
peak
pressure
n = Channel number
< - - >
PCATRIGn
Requests the start of
a PCA trigger
n = Channel number
- - >
Table B-1: ASCII Code Commands used in the Computer Control Protocol