Engineering Tests

Using the arrow keys, move the cursor to the required

selection and press ENTER.

Test Printer

Checks the link to the printer by sending the serial

number, software version and the three lines of report
header to the printer.

Print Settings

Prints the calibration factors for each transducer fitted.

Test Trig Box

When entered, 2 rows of 4 zero’s are displayed. The top

row monitors the trigger buttons on the PCA Trigger /
Remote Call Interface
. The bottom row monitors the
remote call inputs on the PCA Trigger / Remote Call
Interface
. When any of these are triggered, the
appropriate ‘0’ will change to a ‘1’. The left most digit
represents channel 1 and the right most digit channel 4.

Choosing ESC and pressing ENTER will default to the

UTILITIES Menu screen.

Test Parameters

Pressure Units

To cycle the pressure units between mmHg, kPa and
psi, use the up and down keys to change then press
ENTER to accept. The chosen units of pressure
measurement are stored thereafter (until changed via
this screen) and appear on all screens where pressure is
indicated except on screens where back-pressure is
indicated. This is displayed in mmHg for reasons of
high resolution.

Graph Deviation

This sets the upper and lower flow rate deviation lines
on the graphs for Flow Rate and Dual Flow. If this is set
to 0 (Zero) then a single line will be displayed at the set
flow rate. If a flow rate is not entered when a test is set-
up then no deviation lines will be displayed.

PCA Pre Trig Time

When performing Lockout Time tests with PCA pumps,
this sets the time prior to the expiry of the Lock-out
time that the IDA-4 Plus will start attempting to trigger
the pump.

E N G I N E E R I N G T E S T S

T e s t P r i n t e r

P r i n t S e t t i n g s

T e s t T r i g B o x

ESC

T E S T P A R A M E T E R S

PCA

Pressure
Units

Graph

Deviation

Pre Trig

Time

psi

10%

200 sec

Use the up and down arrow

Keys to change.

P r e s s E N T E R t o A c c e p t