Flow test screen (end of test mode), Channel 1 flow, Volume 0.00 – Fluke Biomedical IDA-4 Plus User Manual

Page 39

Advertising
background image

Infusion Pump Analyzer

Operating Instructions

3

3-17

Flow Test Screen (End of Test Mode)

This view of the FLOW TEST screen allows the
operator to save, print or delete the results of the test.

Save

This saves the results of the flow test, or series of tests,
to non-volatile memory for later printing or transfer to a
computer via the RS232 port.

Save & Print

This saves the results as described above, then sends a
report to a printer connected to the printer port on the
instrument. The report will be in either full or summary
format, as previously selected by the operator from the
UTILITIES / PRINTER menu.

Print

This sends a report to a printer connected to the printer
port on the instrument. The report will be in either full
or summary format, as previously selected by the
operator from the UTILITIES / PRINTER menu.

Graph

This switches to the FLOW GRAPH screen for the
current channel to allow a graph of the results to be
viewed before they are saved or printed.

Status

This switches to the STATUS ALL CHANNELS
screen to allow the operator to use another channel or
change settings via the UTILITIES menu.

Delete

This deletes the current set of results from the
operational memory of the instrument, without saving
to the non-volatile memory. CAUTION: Once this
option has been invoked there will not be any way to
recover, view or print the results of this test.

CHANNEL 1

FLOW

0.00

Avg
ml /h

VOLUME 0.00


ml

ELAPSED TIME

00: 00: 00

SAVE

SAVE & PRINT

PRINT

GRAPH

STATUS

DELETE