Flow 2, Volume – Fluke Biomedical IDA-4 Plus User Manual

Page 55

Advertising
background image

Infusion Pump Analyzer

Operating Instructions

3

3-33

Dual Rate Test Screen (End of Test Mode)

This view of the DUAL FLOW TEST screen allows
the operator to save, print or discard the results.

Save

This saves the results of the dual flow test to non-
volatile memory for later printing or transfer to a
computer via the RS232 port.

Save & Print

This saves the results as described above, then sends a
report to a printer connected to the printer port on the
instrument. The report will be in either full or summary
format, as previously selected by the operator from the
UTILITIES / PRINTER menu.

Print

This sends a report to a printer connected to the printer
port on the instrument. The report will be in either full
or summary format, as previously selected by the
operator from the UTILITIES / PRINTER menu.

Status

This switches to the STATUS ALL CHANNELS
screen to allow the operator to use another channel or
change settings via the UTILITIES menu.

Graph

This switches to the DUAL FLOW GRAPH screen for
the current channel to allow a graph of the results to be
viewed before they are saved or printed.

Delete

This deletes the current set of results from memory
without saving to non-volatile memory. Once this
option has been invoked there will not be any way to
recover, view or print the results of this test.

C H A N N E L 1 D U A L F L O W

FLOW 1

0. 00

VOLUME

0. 00

FLOW 2

0.00

ml /h

VOLUME

ml

E l a p s e d T i m e

00: 00: 00

SAVE

SAVE & PRINT

PRINT

GRAPH

STATUS

DELETE

Advertising