Infusion Pump Analyzer

Operating Instructions

3

3-13

Sub Menu1

This menu is reached when all the mandatory fields
have been entered. From this point is possible to go
directly to a test or request further information. Each
menu option is described below. The options on this
menu can be selected by using the arrow keys followed
by the ENTER key.

More

Selecting this option advances to the Manufacturer
field.

Flow

This advances to the FLOW TEST screen where the
operator will be guided through a flow rate test.

Occlusion

This advances to the OCCLUSION TEST screen
where the operator will be prompted through the
occlusion test.

PCA

This advances to the PCA TEST INFORMATION
screen to request information required to perform a
PCA test.

Dual

This advances to the DUAL FLOW RATE
INFORMATION
screen.

Manufacturer

This field is optional and is used to record the
manufacturer of the device under test. This field has a
maximum size of sixteen characters.

Pressing ENTER from this field advances to the Device
Type
field. Pressing ESC steps back to the Flow Rate
field.

Device Type

This optional field is used to record the type number (or
model number) of the infusion device under test. It has
a maximum length of sixteen characters.

Pressing ENTER from this field advances to the Serial
Num
field. Pressing ESC back tracks to the
Manufacturer field.

Serial Num

This optional field is used to record the serial number of
the infusion device under test. A maximum of sixteen
characters is allowed.

Pressing ENTER from this field advances to the
Location field. Pressing ESC goes back to the Device
Type
field.

D e v i c e I n f o r m a t i o n C h a n 1

Control No

12345

Operator Your

Name

Flow Rate

10

Vol 20

MORE

FLOW OCCL PCA DUAL

Manufacturer
Device Type
Serial Num.
Location


P r e s s E N T E R f o r n e x t f i e l d

P r e s s E S C f o r p r e v i o u s f i e l d