18,5 kpa, Sv mv – Infloor Brass Manifold User Manual

Page 6

Advertising
background image

Example of preregulating the valve
Suppose that we need to balance three circuits that have the same pressure loss and loop flow characteristics shown in example

(1.2)

:

Since circuit 2 is the most disadvantaged because it experienced the greatest pressure loss in the panel piping, the remaining circuits must be
adjusted as follows:

Circuit 2

Circuit 1

Circuit 3

ΔP

Loop

= 15 kPa

ΔP

Loop

= 10 kPa

ΔP

Loop

= 7 kPa

G2 = 200 l/h

G1 = 120 l/h

G3 = 80 l/h

ΔP

MV

= 200

2

/115

2

= 3 kPa

ΔP

SV

= 200

2

/287

2

= 0,5 kPa

ΔP

SV

=120

2

/287

2

= 0,2 kPa

ΔP

SV

= 80

2

/287

2

= 0,1 kPa

With the relationship

(1.4)

:

with the relationship

(1.3)

:

with the relationship

(1.3)

:

ΔP

Circuit

= 3 + 15 + 0,5 = 18,5 kPa

ΔP

Circuit

= 10 + 0,2 = 10,2 kPa

ΔP

Circuit

= 7 + 0,1 = 7,1 kPa

+ disadvantaged

H

Predetermined

ΔP

Circuit

= 18,5 kPa

+ disadvantaged

To adjust circuits 1 and 3, we need the
following information to determine the
adjustment position of the micrometric
valves:

Circuit 1
ΔP

MV1

= 8,3 kPa

G1 = 120 l/h

Adjustment position ~ 5.5

Circuit 2
Adjustment position completely open

Circuit 3
ΔP

MV3

= 11,4 kPa

G3 = 80 l/h

Adjustment position ~ 3,5

ΔP

Circuit +

disadvantaged

ΔP

1

ΔP

MV1

ΔP

3

ΔP

MV2

ΔP

2

H

Predetermined

20

10

12

14

5000

1000

100

100

Q (l/h)

ΔP (mm w.g.)

200

50

500

2000

500

16

18

25

30

35

40

45

50

60

70

80

90

120

140

160

180

200

250

300

350

400

450

900

800
700
600

60

70

80

90

120

140

160

180

250

300

350

400

450

1200

1400

1600

1800

2500

3000

3500

4000

4500

50

10

1

2

0,5

5

9

8
7
6

0,6

0,7

0,8

0,9

1,2

1,4

1,6

1,8

2,5

3

3,5

4

4,5

12

14

16

18

25

30

35

40

45

20

ΔP (kPa)

10

0

8

6

4

2

10

0

8

6

4

2

4

5

6

7

8

9

10

1,5

3

2

1

Use of the micrometric balancing valve

The micrometric balancing valves balance each individual circuit in the panels so that the actual design flow is obtained in each one.
Each individual circuit consists of a micrometric balancing valve, panel piping and shut-off valve. The following information must be taken into
account in order to calibrate the system correctly:

· The flow of fluid that must pass through each circuit (design data).

· The pressure loss that occurs in each circuit in accordance with the flow:

ΔP

Circuit

= ΔP

Loop

+ ΔP

SV

(1.3)

· The available head on the panel circuit or predetermined head:

H

Predetermined

≥ ΔP

Circuit

= ΔP

MV

+ ΔP

Loop

+ ΔP

SV

(1.4)

+ disadvantaged

In accordance with the passage of the flow G

Loop

the

micrometric valve must ensure an additional pressure loss
in all the circuits equal to the difference, indicated as ΔP

MV

(Δp micrometric valve).

To allow for an eventual increase in flow, the micrometric
valve of the circuit with the greatest pressure loss may
sometimes be considered as 80% open.

Once the two pieces of information, ΔP

MV

and G

Loop

,

are known for each circuit, the optimal adjustment curve
corresponding to the adjustment position of the valve must
be chosen from the graph.

H

Predetermined

≥ ΔP

Circuit

+ disadvantaged

ΔP

MV

ΔP

Circuit

ΔP

Tot

SV

MV

Advertising