Ivoclar Vivadent Astropol User Manual

Page 3

Advertising
background image

clothing and gloves should be worn.

– Mechanically clean the instruments and immerse them in a

disinfectant solution.

– Sterilize the polishers in the steam autoclave at

134 °C / 275 °F.

– The products are not suitable for sterilization in the chemi-

clave or steam disinfector.

– Care needs to be taken that the selected disinfectant is suit-

able for the disinfection of the polishers and that the instru-
ments are disinfected correctly according to the instructions
of the manufacturer. Please refer to the instructions of the
manufacturer for details on recommended use, immersion
time and suitability of disinfection and cleaning agents.
Aggressive substances, excessively long immersion times or
incorrect dosage may damage the instruments and reduce
the number of re-uses.

Warning
The suitability of the selected method for the disinfec-
tion/sterilization of the respective instrument must be
verified in advance.

Recommendations to prolong the life of polishing
instruments
Mechanical cleaning of the polishing instruments under running
tap water and the subsequent sterilization in the autoclave
increases the number of times the polishing instruments can be
re-used. Alcohol-based disinfectant solutions may reduce
reusability.

Storage and shelf life
– Storage temperature 2 - 28 °C/36 - 82 °F
– Expiry date: See imprint on label

Sterilizable in the steam sterilizer (autoclave) at the

temperature indicated

The respective Instructions for Use can be downloaded

from the Ivoclar Vivadent website.

Keep out of the reach of children!
For use in dentistry only!

Date of preparation: 9/2010; Rev. 2

Manufacturer
Ivoclar Vivadent AG,
FL-9494 Schaan/Liechtenstein

The material has been developed solely for use in dentistry. Processing should be
carried out strictly according to the Instructions for Use. Liability cannot be
accepted for damages resulting from failure to observe the Instructions or the
stipulated area of application. The user is responsible for testing the products for
their suitability and use for any purpose not explicitly stated in the Instructions.

134 °C

Astropol®_GI_WE3_ARTNR_REV_7er_OptraGate-Gi-WE3_590874_Rev2.qxd 01.06.11 14:31 Seite 3