INSTRUCTION MANUAL FOR K-PATENTS PR-01-S (-AX/FM/CS)

DOCUMENT/REVISION No. INM 1/14

Effective: May 15, 2009

30

If a float value is too large for its field it is output in raw decimal floating point format: +1234567+12. The
first sign is the sign of the value. The 7 digit field is the mantissa value after the decimal point. The second
sign is the exponent sign. The 2 digit field is the decimal exponent. The example value +1234567+12 is to
be interpreted as 0.1234567E12, ie. 123456700000.

TEST: The raw refraction value is the same as in the TEST display. The value is displayed with 1 decimal.
The range is 8.0 - 248.0.

Concentration: The concentration value is the same as in the CONC% display. The number of decimals is
controlled by the display decimal parameter in EPROM. The range is dependent of the concentration
factors in EPROM. The value is damped. It is held during wash, hold or recovery phases.

Temperature: The temperature value is the same as in the °C display. The value is displayed with 1
decimal.

Operation phase: For program version 4.0 and newer versions the phase codes are:

Code

Phase codes

0 Blank
1 Wait
2 Precondition
3 Precondition
4 Wash
5 Recovery
6 Hold

Divert status: The divert status field reports the divert control status in coded form, see Section 10.2.

The codes are:

Code Status

0

divert not activated

1 divert

failure

2

channel A, not operating

3

channel B, not operating

4

channel A, operating

5

channel B, operating