INSTRUCTION MANUAL FOR K-PATENTS PR-01-S (-AX/FM/CS)

DOCUMENT/REVISION No. INM 1/14

Effective: May 15, 2009

87

WASH

STOP ?

EXTERNAL

HOLD ?

Start
prism
wash

REMOTE WASH

SWITCH

NO

YES

YES

NO

Automatic Wash start

1. Wash interval timer
2. Smart Wash

YES

NO

PRECONDITIONING

WASH

RECOVERY

WASH

CHECK

OK?

RETRY

WASH ?

NO

YES

“Prism wash

failure“

Reset “Prism

wash failure“

Manual

Figure 9.41

Wash decision logics.

WASH STOP LOGICS:

1. The preconditioning and wash relays are not activated by automatic wash start under "Low conc/no

sample", Section 6.30, as this indicates a clean prism in an empty process line.

2. Wash can be prevented when the process temperature is below a limit. The Wash stop function is used

to prevent steam wash when the process pipe is empty or when the process is stopped. To activate the
"wash stop" key sequence Calibrate/Parameters/1. Prism wash/Wash stop/1. Activate and set the
temperature limit. The default value for wash stop is inactive. When the wash is not accepted the
diagnostics message is "Wash stop/temp. limit", see Section 6.3.

3. Wash can also be prevented by an external contact closure, telling e.g. that the process pump has

stopped. See Section 2.8, paragraph 3.3.