K-Patents PR-01-S User Manual

Page 6

Advertising
background image

INSTRUCTION MANUAL FOR K-PATENTS PR-01-S (-AX/FM/CS)

DOCUMENT/REVISION No. INM 1/14

Effective: May 15, 2009

4

2.2. STANDARD SPECIFICATIONS

Refractive Index ranges:

Low range R.I. 1.320...1.460
High range R.I. 1.380...1.530

Max. span:

R.I. 0.08 (See Section 5.6)

Accuracy:

R.I. ± 0.0002 (corresponds typically to ± 0.1% by weight). Repeatability and
stability correspond to accuracy.

Speed of response:

0.8 s undamped

Damping time constant:

Selectable up to 5 min

Process temperature:

max. 150°C (300°F) (for higher temp. consult factory)

Ambient temperature:

Sensor: max. 45°C (113°F), min. -20°C (-4°F),

Indicating transmitter: max. 50°C (122°F), min. 0°C (32°F)

Process pressure:

Flange connections up to 25 bar (350 psi)
Sanitary clamp max. 15 bar (200 psi) at 20 °C (70 °F)/9 bar (125 psi) at 120
°C (250 °F)

Recommended flow velocity:

above 1.5 m/s (5 ft/s)

Wetted parts:

AISI 316L stainless steel, prism gaskets Kalrez (prism pads teflon)

Sensor weight:

Sandvik-clamp 7 kg (15 lbs)/ Flange DIN/ANSI/JIS 10.5 kg (23 lbs)

Process connection:

NS 80 (3" Sch 10s) (Sandvik clamp/Flange DIN/ANSI/JIS/Sanitary clamp)

Current output

4-20 mA/0-20 mA, max. load 1000 Ohm, Galvanic isolation 1500 V DC or
AC (peak), Built-in hold function during prism wash.

Serial output:

RS485/RS232
Galvanic isolation 500 V DC or AC (peak)

Power:

100-115 V/220-240 V, 50/60 Hz, 15 VA

Interconnecting cable:

Shielded cable, 2 twisted pairs with individual shields, 0.5 mm

2

.

Digital transmission according to RS485.

Interconnecting cable length:

Standard 10 m (33 ft), max. 100 m (330 ft)

Indicating transmitter:

Enclosure IP66 Nema 4X

Indicating transmitter weight:

4.5 kg (10 lbs)

Alarms:

Two built-in signal relays, max 24V, 500 mA

Options:

- Long probe version (-LPL). Probe length is 152 mm (6") longer than
standard.
- Wetted parts: Hastelloy C, titanium or palladium doped titanium
- Cable fittings to the Indicating transmitter: European cable glands or
M20x1.5 metric cable glands or US conduit hubs, see Section 7.3.
- Built-on prism wash nozzle, see Section 8.3.
- Intrinsically Safe PR-01-S-AX/FM/CS, see Section 12. ATEX EX II 1 G /
Cenelec Eex ia IIC T4;

Factory Mutual (FM) approved Class I, Div. 1,

Groups A, B, C, D T4 (T

amb

=45 °C); CSA certified for Class I, Groups A-D

Accessories:

- External output unit, Section 10.1
- Digital Divert Control Unit DD-01, Section 10.2
- Retractor and isolation valve, Chapter 11.
- Prism wash nozzles for steam and hot water, Section 3.4.
- Flow cells according to separate drawings available from K-Patents.

(We reserve right to technical alterations.)

Ordering information:
- Desired scale, properties of process solution
- Process temperature and pressure range
- Process flow range and pipe diameter
- Desired process connection


- Length of interconnecting cable
- Supply voltage and frequency
- Options and accessories
- User tag