Figure 5.60

Sensor rangeability.

5.7. EPROM PARAMETERS

How to enter the calibration parameters into the non-volatile EPROM memory is described in Section 2.8. The
factory settings are found in the DELIVERY DATA SHEET. Figure 5.70 shows how the microprocessor
program is using the parameters.

BIAS
GAIN
CUBE
SQUARE

TNM

TCOEFF
TEMP TCADJ.
CONC TCADJ.

SQUARE TEMPCO

DAMPING

TEST

TEMP

UNIT

DECIMALS

ZERO
SPAN

HOLD

UNIT
°C / °F

CONC.

DISPLAY

OUTPUT
mA

TMPC 0
TMPC 1

RI-BIAS

RI-GAIN

RI-CUBE
RI-SQUARE

STAN-
DARD
RI

DISPLAY

FIXED

RI
TEMP.
COMP.

TBIAS

TEMP.
DISPLAY

TEMP °C

DATA

SENSOR

ADAPTION

Figure 5.70

Calculation flow diagram.

The ENTER procedure contains a check of the EPROM parameters format. A number outside the range
limits cannot be entered. Only if the range limits of a parameter contain a decimal point, the parameter may
contain a decimal point. For some parameters the number of characters is limited (character = digit,
decimal point or minus sign).

Indicating transmitter exchange:

If the Indicating transmitter is exchanged, then all parameters and configuration determinations have to be
changed.