System timer overview – Quantum Composers 9500 Plus Series User Manual

Page 32

Advertising
background image

28

Edge:

Press the NEXT key until the Edge parameter is high-
lighted. Set the instrument to trigger off the rising edge
or falling edge as desired.

Pulse Width, Delay:

Enter the Channel menus by pressing the letter key.

Enter the required pulse width and delay. Repeat for
each output channel.

Start:

Press the RUN/STOP key to start/arm the instrument.

The 9500+ will now generate a pulse for every valid
trigger.

Stop:

Press the RUN/STOP key a second time to stop/dis-
arm the instrument (i.e. to stop generating pulses).

System Timer Overview

For internal operation, the 9500+ contains a timer and mode generator which

generates an internal T

o

clock that is used to trigger all the channel timers.

System modes are controlled via the MODE menu.

To Use Continuous Mode

The RUN/STOP button starts and stops a continuous pulse stream at the rate
specified by the Rate menu. This corresponds to the normal output mode for
most pulse generators. To generate a continuous stream of pulses:

- Within the system MODE menu

Mode:

Select Continuous for the system mode.

- Within the RATE menu

Period:

Set the desired period.

Pressing the RUN/STOP key will now generate a stream of T

0

pulses, at a rate

specified by the period parameter

To Use Single Shot Mode

To generate a single pulse with every press of the RUN/STOP key:

- Within the system Mode menu

Mode:

Select Single Shot for the system mode.

Pressing the RUN/STOP key will now generate a single pulse.