T Option

System Period

The minimum period is 50 ns (20 MHz max frequency).
The maximum period is 10 s (.1 Hz frequency).

Pulse Range Limitations

The maximum pulse width is 5 s.
The maximum pulse delay is 5 s

Multiplexer

Multiplexing differs from the base model as follows:

1. Multiplexing is implemented in channel pairs. Channel A and B are paired,
Channel C and D, etc. The enable and disable parameter for each channel
turns on and off inclusion of the internal pulse of the other channel in the pari for
the current channel’s output.

2. The SCPI command for the multiplex parameter use a string “ENABle” or
“DISable”. For example, “:PULS1:MUX ENAB\n” causes Channel B’s internal
pulse to be present on Channel A’s output.

Channel Menus

Timing Parameters

Pulse Configurations

Channel Mode

Gate/Mux

Enable

Enable

Enable

Enable

Sync Source

Polarity

Mode

Max Enable

Pulsewidth

Output Type

On Cycle

Gate Enable

Delay

Output Level

Off Cycle / Burst

Wait



Clock In/Clock Out

The “T” option pulse generator has Clock In and Clock Out functionality to allow
synchronization of several instruments to an industry standard 10 MHz signal.
The rear panel of the unit has BNC connectors for Clock In and Clock Out
signals.

The “Clock In” BNC connector allows synchronization of the unit’s timers to a
signal generated externally. When the Clock In is enabled, a 10 MHz pulse must
be present on the Clock in BNC for the unit to generate pulses.