46

Keyword

Lower-Level

Keyword

Parameter

Std /

New

Comments

:DISPlay

Std

Subsystem. Contains commands to control the display.

:MODe

0 / 1 or OFF / ON

New

Enables / Disables automatic display update. When

true, front panel display is updated with serial command

parameter changes. Setting to false (0) decreases

response time.

:UPDate?

New

Query only. Forces update of display. Use when mode

is false.

:BRIGhtness

0 - 4

New

Controls intensity of display. Range is 0 to 4, where 0 is

off and 4 is full intensity.

NOTE: DISPlay commands are considered irrelevant for the 8510 series.







IEEE 488.2 Common Commands

Mnemonic

Command Name

Parameter

Comments

*IDN?

Identification Query

Queries the pulse generator identification. The ID will be in

the following format: model# - #channels - option# -

version#

*RCL

Recall Command

0 - 12

Restores the state of the pulse generator from a copy stored

in local nonvolatile memory (0 through 12 are valid

memory blocks).

*RST

Reset Command

Resets the pulse generator to the default state.

*SAV

Save Command

1 - 12

Stores the current state of the pulse generator in local

nonvolatile memory (1 through 12 are valid memory

blocks).

*TRG

Trigger

Generates a software trigger pulse. Operation is the same as

receiving an external trigger pulse.

*LBL

Setup Label

? or "string value"

The <?> is used to query the label of the last saved or

recalled configuration.

The string must be in double quotes and no longer than 14

characters. Command must be followed by a *sav [1 / 2 / n]
command to take effect. Note: to see the label on the screen

a display update or reboot must take place