9500+ SCPI Command Summary

Keyword

Lower-Level

Keyword

Lower-Level

Keyword

Lower-Level

Keyword

Parameter

Std /

New

Comments

:SYSTem

Std

:STATe?

New

Query Only. Returns the state of

the unit: Returns "1" if the unit is

armed and/or generating pulses

or "0" if the unit has been

disarmed

:BEEPer

Std

Subsystem. Controls the audible

beeper.

:STATe

0 / 1 or OFF / ON

Std

Enables / disables the beeper.

:VOLume

0 - 100

Std

Sets the volume of the beeper.

Range is 0 to 100, where 0 is off

and 100 is maximum volume.

:COMMunicate

Std

Subsystem. Controls the RS232

and GPIB interfaces.

:GPIB

Std

Subsystem. Controls the

physical configuration of the

GPIB port.

:ADDRess

1 - 15

Std

Sets the GPIB of the instrument.

:SERial

Std

Subsystem. Controls the

physical configuration of the

RS232 port

:BAUD

4800 / 9600 /

19200 / 38400 /
57600 / 115200

Std

Sets the baud rate for both

receiving and transmitting using

the DB9 RS232 port.

:USB

4800 / 9600 /

19200 / 38400 /
57600 / 115200

New

Sets the baud rate for

communication when using

mapped comports for USB and

Ethernet communication. Must

be set to the default valve

(115200) for raw TCP/IP

communication (i.e. Labview).

:ECHo

0 / 1 or OFF / ON

New

Enables / Disables transmission

of characters received on the

DB9 serial port.

:KLOCk

0 / 1 or OFF / ON

New

Locks the keypad.

:AUTorun

0 / 1 or OFF / ON

New

After power-up, the unit will start

generating pulses automatically.

:VERSion?

New

Query only. Returns SCPI

version number in the form:

YYYY.V (ex. 1999.0)

:SERNumber?

Std

Query only. Returns the serial

number of the unit.











Advertising