Quantum Composers 9250 Series User Manual
Page 38

9250 Manual Version 1.2 | Quantum Composers
38
:PERiod
40[ns]
– 4000[s]
Sets the T
0
period. The
command should be sent
without units. If for example
200ns is desired the
parameter sent should be
200e-9, or the decimal
equivalent.
:MODe
NORMal/SINGle/
BURSt/DCYCle
Changes the system output
mode.
:BCOunter
1-1,000,000
Changes the number of
pulses to output when the
system is in burst mode.
*Note: The commas should
be omitted.
:PCOunter
1-1,000,000
Changes the number of on
pulses to output when the
system is in Duty Cycle
mode.
*Note: The commas should
be omitted.
:OCOunter
1-1,000,000
Changes the number of off
pulses to suppress when the
system is in Duty Cycle
mode.
*Note: The commas should
be omitted.
:CCOunter
0-1,000,000
Cycle counter. Sets the
number of system cycles to
repeat when in duty cycle
mode. Setting to zero
disables any cycle counts.
:EXTernal
Submenu for selecting the
trigger input
:MODe
DISabled /
TRIGger /
GATe
Selects the trigger mode*
:LEVel
0.2 - 15[V]
Choose the gate level
threshold to trigger on which
should be set to ~ 50% of the
input voltage.
:EDGe
RISing / FALLing
Selects which edge (rising or
falling) to use as the trigger
signal
:POLarity
LOW / HIGH
Sets the polarity of the gate
signal. HIGH output is active
when the gate signal is high;
LOW output is active when
the gate signal is low.
:EXTPeriod
External trigger period.
:STATe
0/1 or OFF/ON
Enables/Disables the counter
function.
:AUTo
0/1 or OFF/ON
Sets auto update mode. Auto
will continually update the
value, overwriting the
previous value.
:CLear
Clears the active counter.