Appendix a – specifications, Pecifications, 9250 specifications – Quantum Composers 9250 Series User Manual

Page 42

Advertising
background image

9250 Manual Version 1.2 | Quantum Composers

42

9. Appendix A – Specifications

9250 Specifications

Indoor use in dry conditions only

Ordinary Protection: This product is NOT protected against harmful ingress of moisture.

Class III Equipment (external 5VDC power source)

External Power Supply: Class II (ground connection not required for safety)

o Mains Supply Input: 100-240V ~ 50-60Hz 0.3A
o Output: 5VDC 2A

Pollution Degree II (Micro-Ambient Pollution restricted to temporary conductivity caused

by condensation).

External Power Supply: Installation (Overvoltage) Category II for transient over-voltages.

Maximum Relative Humidity: 0-80% relative humidity, non-condensing

Maximum altitude: 0 to 2000m (6562 ft.)

Accessories: USB cable & 5V (10W) DC Power Supply

Cleaning Instructions: Light dusting with cloth damp with water and/or usage of

compressed air is all that is needed.



9250 Specifications

MIN

TYP

MAX

UNIT

I/O Configuration

Model/Output

9254

– 4 Independent + 4 Virtual Channels (Mux)

Input

1 Selectable Trigger/Gate Input

Internal Rate Generator

Rate (T

0

Period)

0.00025

-

25,000,000

Hz

Resolution

-

4

-

ns

Accuracy

4ns + (0.0001 x Period)

T

0

Period Jitter

-

-

40

ps(RMS)

Time Base

250MHz, Low Jitter PLL

Oscillator

25MHz, 280ppb Crystal Oscillator

System Modes

Single, Continuous, Burst, Duty Cycle

Burst Mode

1

-

1,000,000

Pulses

Duty Cycle Mode

1

-

1,000,000

Pulses

Pulse Control Modes

Internal Rate Generator, External Trigger/Gate