Chapter 2 Front Panel Operations

RIGOL

DSG3000 User’s Guide

2-3

Phase Offset

Set the phase offset of the RF signal.

When multiple RF signal generators are used to output signals, you can adjust the

phase offset of the instrument to synchronize the phase or set a fixed phase offset

for the signals.

Press FREQPhase Offset, after that use the numeric keyboard to input the
value of the phase offset and press deg or the unit key.
 Press Enter to select the default unit (deg).
 Press the left/right direction keys to enter the parameter editing state.
 You can also use the up/down direction keys or knob to modify the phase offset

at the current step.

 Press FREQ Phase Offset and then press Step to set the step.
 Press Rst Phase to reset the current phase offset to 0 deg.

Explanation

1. When the frequency offset is 0 Hz, the frequency displayed in the interface

(namely setting frequency) is equal to the actual output frequency.

2. When frequency offset is not 0 Hz, the output frequency is determined by

setting frequency and frequency offset. The three parameters satisfy the

equation: setting frequency (display frequency) = output frequency +

frequency offset

3. The deviation (namely the actual output frequency) between interface display

frequency and frequency offset cannot exceed the frequency range of

instrument.