RIGOL DSG3000 Series User Manual

Page 83

Advertising
background image

Chapter 2 Front Panel Operations

RIGOL

DSG3000 User’s Guide

2-43

Bus Trigger

After selecting “Bus” trigger, if the sweep mode is set to “Cont”, the

instrument starts a sweep each time the “*TRG” command is sent. If the

sweep mode is “Single”, press Single to meet the single sweep condition.

After that the instrument starts a sweep and then stops each time the

“*TRG” command is sent.

External Trigger

In external trigger, the RF signal generator accepts the trigger signal input

from the [TRRIGER IN] connector (as shown in the figure below) at the

rear panel. If the sweep mode is set to “Cont”, the instrument starts a

sweep each time a TTL pulse signal with specified polarity is received. If the

sweep mode is “Single”, press Single to meet the single sweep condition.

After that the instrument starts a sweep and then stops each time a TTL

pulse signal with specified polarity is received.

To specify the polarity of the TTL pulse signal, press Trig Slope to select

“Pos” or “Neg” and the default is “Pos”.

Advertising