Amprobe, Measuring procedures, Caution – Amprobe DMIII-Multitest Power-Quality-Recorder User Manual

Page 65

Advertising
background image

AMPROBE

DMIII MULTITEST

EN-63

8. MEASURING PROCEDURES

8.1.

USING THE INSTRUMENT IN A SINGLE-PHASE SYSTEM

CAUTION

The maximum voltage between V1 and COM inputs is 600 V~ (CATII) /
350V~ phase

– earth or 600V~ (CATIII) / 300 V~ phase to earth.

Do not measure voltages exceeding the limits prescribed by this manual.
Should you exceed the voltage limits you could damage the instrument
and/or its components or endanger your safety.

Phase

Neutral

I1

T1 T2 T3 T4

I1 I2 I3

(V1) (V2) (V3)

(COM)

Instrument connection in a single-phase system

1. Check, and if needed modify, the basic settings of the instrument (see paragraphs 7.1

and 7.2). Particularly, the single-phase mode must be set.

2. Rotate the switch to the position corresponding to the type of analysis desired.
3. Connect the phase and neutral voltage wires respecting the connections shown in the

picture.

4. If you want to measure current and power, connect the clamp meter to the phase

conductor respecting the specifications shown on the clamp and the connections
shown in the picture.
In case of doubt, select the position POWER and check if the active power P is
positive. If it’s negative, remove current transducer from the wire and reconnect it so
the transducer label faces the opposite direction.

5. Apply voltage to the electrical equipment under test (if previously shut off for the

instrument connection).

6. The values of the available electrical parameters will be displayed on the display of the

instrument. For further details see the paragraph relevant to the position of the switch.

7. You can press HOLD to interrupt the updating in real time of the displayed values.
8. If you want to record:

a) Check, and if needed modify, the values of the basic parameters (see paragraphs

7.1 and 7.2).

b) Check, and if needed modify, the recording parameters by pressing MENU (see the

paragraph corresponding to the position of the rotary switch selected).

c) To start the recording press START (see paragraph 10.1).


Advertising