Optimizing pump therapy 85

5

The SQUARE DURATION screen will appear. Enter the amount of time
you want the Square Wave bolus to last and press

ACT

.

NOTE -

If you have the BG Reminder feature turned on, a screen

displays allowing you to accept or modify the length of time after
this bolus before you are reminded to check your blood glucose.

6

The

BOLUS DELIVERY screen will appear. The pump will beep or vibrate at the start and end of the

bolus. As the bolus is delivered, the screen will show the bolus type and amount until the total units
have been delivered.

For a

Square Wave bolus

do these steps:

For a

Dual Wave bolus

do these steps:

a.

Select

Square Wave Bolus

. Press

ACT

.

The SET SQUARE BOLUS screen will appear.

b.

Enter the desired amount
for the Square Wave
bolus units and press

ACT

.

c.

Continue to step

5

.

a.

Select Dual Wave Bolus and press ACT. The SET
DUAL BOLUS TOTAL screen will appear.

b.

Enter the desired amount
for the total dual bolus
units. Press

ACT

.

NOTE -

The number of units you enter for

the SET DUAL BOLUS TOTAL is a total of

both the normal and Square Wave bolus

units.

c.

In the next screen, press

/

to change the

normal (Now) and Square
portions of the Dual Wave
bolus. Press

ACT

. Notice

the screen also shows the
percentage amount for each portion.

d.

Continue to step 5.

SET SQUARE BOLUS

1.2U

SET DUAL BOLUS TOTAL

1.2U

0.6U

0.6U

Now:
50%
Square:
50%

ACT

SQUARE DURATION
Sq: 1.2U

0:30