Default (factory) settings – Energizer 715 User Manual

Page 176

Advertising
background image

154 Chapter: 11

Default (factory) settings

#6: Previous bolus activity results in a calculation of 3.5 units unabsorbed (active) insulin. User inputs 60 grams of carbs

and a blood glucose of 200.

Menu

Item

Default Setting Limits

Increments

Warning

Limits

Bolus
Menu:

Bolus Wizard feature: Off

Easy bolus: On

Easy bolus step: 0.1 u/h

maximum bolus setting

Dual/Square bolus: Off

Maximum bolus:

10.0 u/h

0 - 25 u (per single
bolus)

BG reminder: Off

0:00 - 5:00

0:30 (minutes)

Basal
Menu:

Patterns: Off

Maximum basal rate: 2.0 u/h

0.00 - 35.00 u/h

0.05u

Basal rate: 0.0 u/h

0.05u

Temp basal type: U/H

max basal rate setting

0.05U/H (or
5%)

(basal rate =
0.0)

60g

30g/u

200mg/dL - 120mg/dL

40mg/dL/u

* This negative number indicates that active insulin is
sufficient to cover the correction that is needed. Thus,
correction will be 0 units. Active insulin does not reduce
the food portion of the estimate.

(correction)

= -1.5 units*

- 3.5 units (active insulin)

= 2 - 3.5

(food estimate)

= 2 units

= 2 + 0

= 2 units

estimate = 2 units

+