18 chapter: 2 – Energizer 715 User Manual

Page 40

Advertising
background image

18 Chapter: 2

Special

- indicates a special feature is active or an alert condition(s)

exists. Special mode does not restrict any of the pump functions.
When the pump is in Special mode, an open circle appears at the
top of the screen and it will beep/vibrate periodically to remind
you of the condition. The conditions and features that will put the
pump in Special mode are:

Attention

- indicates insulin delivery has stopped. This can mean

that the pump is in Suspend mode. It can also mean an alarm is
active or an alarm condition exists that needs immediate
attention for insulin delivery to resume. A solid circle appears at
the top of the screen and the pump will beep periodically until
the either the pump is taken out of Suspend mode or the condition
is cleared. The screen will show text describing the condition that
put the pump in Attention mode. For example, if the reservoir is
empty, “Empty Reservoir” will appear on the screen.

When the pump is in Attention mode, it will beep/vibrate periodically to remind you of the
condition. The beep/vibrate frequency varies depending on the condition that put the pump in
Attention mode. See “Alarm conditions” on page 137 for alarm conditions that will trigger the
Attention mode. See “Stopping your pump” on page 39 to learn about Suspend mode.

Low Reservoir condition

Low Battery condition

Block feature is on

Dual or Square Wave bolus
delivery

Basal pattern A or B is active

Temporary basal
delivery