Pump specifications 143

Chapter: 11

Pump specifications

This section provides detailed information on specifications related to your pump. The safety features of your
pump are individually listed and described.

Alarms and error messages

➠

Indicators: audible tone (beep) or vibration (silent)

➠

All alarms and errors show messages on the pump’s screen and provide instructions on what to do.

Unresolved alarms will escalate to siren mode for added safety.

Alarm history

➠

The maximum records shown is 36

Backlight

➠

Display type Liquid Crystal Display (LCD)

➠

Time-out period is 30 seconds. The backlight will turn off when you do not push any buttons for 30

seconds or more.

Basal

Maximum patterns

delivery rate

warning limits

3 patterns (with 48 delivery
rates for each pattern)

rate increments:

0.05 units/hour

lowest possible rate:

0.05 units/hour

highest possible rate:

35.0 units/hour

factory maximum rate:

2.0 units/hour

lower limit:

less than 90 mg/dL

upper limit:

greater than 140 mg/dL