The pump delivers basal amounts in 0.05 U/H increments. Because of this, your temp basal

amount will be rounded down to the next 0.05 U/H increment.

temp basal start time

(120% x rate 3 = 2.34)
2.34 rounded down to
the next 0.05 increment
is 2.30 U/H

1.50 U/H

(rate 1)

2.55 U/H

(rate 2)

1.95 U/H

(rate 3)

3.05

U/H

2.30

U/H

(120% x rate 2 = 3.06)

3.06 rounded down to

the next 0.05 increment

is 3.05 U/H

temp basal stop time

temp basal settings

temp basal type: Percent of basal

duration: 4 hours (1:05p - 5:05p)

rate: 120 percent (%)

5:05 p.m.
(17:05)

7:00a.m.