Energizer 715 User Manual

Page 35

Advertising
background image

The basics 13

5

While the pump turns on, it will show various screens until the HOME screen appears.

NOTE -

If you are replacing a battery for a pump that has very recently been in use, the only

screen you will see before the HOME screen is the one displaying the version number.

If the HOME screen does not appear, do these steps:

a.

Check that the battery is inserted correctly. If the battery has been installed backwards,
remove the battery and install it properly.

b.

If your pump still does not turn on or you get a FAILED BATT TEST alarm, remove and replace
the battery with a new one.

c.

If the pump is still not on, call the Medtronic MiniMed 24-hour Product Help Line.

6

Check to make sure the time and date are correct. If this is the first battery installed, or if more
than 5 minutes have passed since you removed the battery, you will be prompted to check the time
and date. See “Setting the time and date” on page 23 for programming instructions.

7

Press

ESC

to view the STATUS screen, making sure no alarms are active. If an alarm is active, follow

the instructions on the screen.

NOTE -

Make sure that you dispose of the battery per your local battery disposal regulations.

7

Ver X.XX

HOME screen

Advertising
This manual is related to the following products:

515