Connecting quick-set disconnecting quick-set – Energizer 715 User Manual
Page 76
54 Chapter: 4
Connecting Quick-set
Disconnecting Quick-set
The Quick-set allows you the freedom to temporarily disconnect from your pump without removing the
infusion set from your body.
If you inserted a new reservoir, you should have already done a manual prime. A manual prime is
always done to fill the infusion set tubing with insulin after you rewind your pump and insert the
reservoir. If you are changing your infusion set without changing your reservoir, do a fixed prime to
fill the tubing. See “Fixed prime” on page 55 for instructions.
Place the connector part (flat-side facing down) on the infusion set until it is fully seated. Do not
grip the connector part by the flat side-grips.
Squeeze the side grips of the connector part
with your fingers.
Remove the connector from the site.
Infusion set connected
Infusion set
Connector part
Side grips