Pump specifications 153

#3: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a blood glucose of 70.

#4: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a blood glucose of 100.

#5: Previous bolus activity results in a calculation of 1.5 units unabsorbed (active) insulin. User inputs 60 grams of carbs

and a blood glucose of 200.

70mg/dL - 90mg/dL

40mg/dL/u

-20mg/dL

40mg/dL

+

= 2 + (-0.5)
= 1.5 unit

estimate = 1.5 unit

(correction)

= -0.5 unit

=

60g

30g/u

(food estimate)

= 2 units

+

Correction is 0 because the current blood glucose
reading is between the BG High and Low targets

(correction)

= 2 + 0

= 2 units

estimate = 2 units

60g

30g/u

(food estimate)

= 2 units

60g

30g/u

200mg/dL - 120mg/dL

40mg/dL/u

(food estimate)

(correction)

= 2 units

= 0.5 units

- 1.5 units (active insulin)

= 2 - 1.5

= 2 + 0.5

= 2.5 units

estimate = 2.5 units

+

Advertising