Utilities 111

NOTE -

If you use “time” as the low resv warning type and you deliver large boluses, the actual

time remaining could be less than the warning time. “Time” low resv warning types are intended
to let you know if you will have enough insulin while you are sleeping.

For “Insulin units”:

Enter the number of units you want remaining
when the first warning will go off. Press

ACT

.

The pump will alarm first when the specified
units remain, then again when half that
remaining amount is used.

For “Time”:
Enter the amount of remaining time you want for
the first warning. Press

ACT

.

The pump will alarm first when the specified time
remains, then again one hour before empty.

(flashing)
can be 5.0 – 50.0

WARNING UNITS

20

U

(flashing)
can be 2 –24

WARNING TIME

8:00

Hours