Is priming complete, Max delivery, Motor error – Energizer 715 User Manual

Page 161: No delivery, No reservoir, Troubleshooting 139 is priming complete

Advertising
background image

Troubleshooting 139

Is priming complete?

If you manually prime your pump with more than 30u insulin, this message

appears. Press

ESC

,

ACT

to clear the message. If manual prime is complete,

press

ESC

. If the manual prime is not complete, press and hold

ACT

until

manual prime is complete.

Max delivery

This alarm alerts you when you have taken more insulin than expected

based on maximum bolus and maximum basal rates.

Motor error

Insulin delivery has stopped. This alarm occurs if your pump detects a

motor error.

No delivery

Insulin delivery has stopped. This alarm occurs if your pump detects a

blockage.

No reservoir

The reservoir is not inserted correctly or no reservoir has been inserted.

IS PRIMING

COMPLETE?

Disconnect from body

for instruction

See user guide

ESC, ACT to clear

MAX

DELIVERY

Exceeded 1 hour

max delivery

Check BG

3:25P

ESC, ACT to clear

MOTOR

ERROR

Delivery stopped
Disconnect set

ESC, ACT to clear

3:25P

Delivery stopped
Change entire set
Check BG

See user guide
to troubleshoot

ESC, ACT to clear

NO

DELIVERY

11:17A

NO

RESERVOIR

Delivery Stopped

Change reservoir set

ESC, ACT to clear

4:36P